FDA Brief

Company record

Pivotal Health Solutions, Inc.

0 FDA warning letters since 2023 and 2 recalls in the last 12 months.

Get an email the day FDA posts anything new about Pivotal Health Solutions, Inc..

Recalls

ReportedClassFirmProduct and reason
Dec 3, 2025 Class II Pivotal Health Solutions, Inc.
Device
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
FDA notice Z-0587-2026
Dec 3, 2025 Class II Pivotal Health Solutions, Inc.
Device
Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
FDA notice Z-0588-2026